生命科学 eQMS 12 项
Depyrogenation and Autoclave Sterilization Validation Checklist
此检查清单为英文版。
Verifies that moist-heat (autoclave) sterilization and dry-heat depyrogenation processes are validated for lethality, endotoxin reduction, and load reproducibility. Use to plan or audit thermal sterilization validation.
USP <1211>USP <1228>ISO 17665EU GMP Annex 1
免费 PDF 或 Word · 输入邮箱即可下载。
Autoclave (Moist Heat) Validation
Establish steam sterilization lethality.
- Are heat penetration and temperature distribution studies performed across each validated load configuration? *
- Is a minimum F0 defined and achieved at the coldest point of the load? *
- Are biological indicators (Geobacillus stearothermophilus) used to demonstrate a minimum 6-log reduction? *
- Are worst-case (maximum and minimum) loads challenged? *
Dry-Heat Depyrogenation Validation
- Is a minimum 3-log reduction of endotoxin demonstrated using endotoxin indicators (spiked at >=1000 EU)? *
- Are heat penetration studies performed to confirm the FH lethality across the tunnel or oven load? *
- Is belt speed, temperature, and airflow controlled and recorded for continuous depyrogenation tunnels? *
Equipment Qualification and Controls
- Are temperature and pressure sensors calibrated with traceable standards before and after studies? *
- Is cycle control instrumentation qualified (IQ/OQ/PQ) for the sterilizer or tunnel? *
- Are alarms and interlocks challenged for out-of-range conditions?
Routine Monitoring and Revalidation
- Is each production cycle monitored against validated critical parameters and reviewed for release? *
- Is periodic revalidation performed and triggered by changes to load, equipment, or process? *
下载完整的Depyrogenation and Autoclave Sterilization Validation Checklist检查清单
免费获取排版清晰、便于打印的 PDF,或可编辑的 Word 文档。